§ 868.1985 - Spinal imaging system for neuraxial procedures.
(a) Identification. A spinal imaging system for neuraxial procedures is a sensor and software system that is used to assist in identification of the interspinous space for neuraxial procedures.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must evaluate:
(i) Rate of neuraxial procedure success across the intended use population;
(ii) Procedure time;
(iii) Adverse events, including pain, bleeding, unintended dural puncture, and injury to other spinal structures; and
(iv) Comparison of performance to manual palpation or alternative method of determination of the target insertion site.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must evaluate the accuracy, repeatability, and resolution of imaging of the underlying anatomy representative of intended use populations.
(3) Human factors testing must demonstrate that the user can correctly use the device based solely on the directions for use.
(4) Software verification, validation, and hazard analysis must be performed. Validation testing must verify and validate proper image construction.
(5) Electromagnetic compatibility and electrical, mechanical, and thermal safety testing must be performed.
(6) The patient-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance testing must demonstrate the sterility of any device components intended to be labeled sterile.
(8) Labeling must include a warning that risks inherent to neuraxial procedures under standard of care conditions are still present with use of the device.