§ 1065.377 - Interference verification for NH3 analyzers.
(a) Scope and frequency. This section describes how to perform interference verification for certain analyzers as described in § 1065.277. Perform interference verification after initial analyzer installation and after major maintenance.
(b) Measurement principles. Certain compounds can positively interfere with analyzers by causing a response similar to NH
(c) System requirements. Analyzers must have combined interference that is within (0.0 ±2.0) µmol/mol.
(d) Procedure. Perform the interference verification as follows:
(1) Start, operate, zero, and span the NH
(2) Except as specified in paragraph (d)(9) of this section, select a multi-component span gas meeting the specification of § 1065.750 that incorporates the all the appropriate interference species. Use a humidity generator that meets the requirements in § 1065.750(a)(6) to humidify the span gas. If the sample does not pass through a dryer during emission testing, humidify your test gas to an H
(3) Introduce the humidified interference test gas into the sample system upstream or downstream of any sample dryer, if one is used during testing.
(4) If the sample does not pass through a dryer during this verification test, measure the H
(5) If the verification procedure does not include a sample dryer, use good engineering judgment to prevent condensation in the transfer lines, fittings, or valves between the point of
(6) Allow time for the analyzer response to stabilize. Stabilization time may include time to purge the transfer line and to account for analyzer response.
(7) Operate the analyzer to measures the sample's NH
(8) The analyzer meets the interference verification if the result of paragraph (d)(7) of this section meets the tolerance in paragraph (c) of this section.
(9) You may instead perform interference verification procedures separately for individual interference species. The interference verification specified in paragraph (c) of this section applies based on the sum of the interference values from separate interference species. If the concentration of any interference species used is higher than the maximum levels expected during testing, you may scale down each observed interference value by multiplying the observed interference value by the ratio of the maximum expected concentration value to the concentration in the span gas. You may run separate H